Astron. Astrophys. 358, 708-716 (2000)
2. Observations and data reduction
Imaging spectrophotometry was performed with the
32 32 element mid-IR camera (CAM) on
board the ISO satellite, using the Circular Variable Filters (CVFs)
(see Cesarsky et al. 1996a for a complete description). The
observations employed the 6" per pixel field-of-view of CAM. Full
scans of the two CVFs in the long-wave channel of the camera were
performed with both increasing and decreasing wavelength. The results
of these two scans are almost identical, showing that the transient
response of the detector was only a minor problem for these
observations. The total wavelength range covered is 5.15 to
16.5 µm and the wavelength resolution
.
10 0.28 s exposures were added for
each step of the CVF, and 7 more at the first step in order to limit
the effect of the transient response of the detectors. The total
observing time was about 1 hour. The raw data were processed as
described in Cesarsky et al. (1996b), with improvements
described by Starck et al. (1998) using the CIA
software 1. The
new transient correction described by Coulais & Abergel (1998) has
been applied but the corrections introduced are minimal, as mentioned
above. The bright star
Ori A is visible in the
maps of several spectral components and has been used to re-position
the data cube. This involved a shift of only 2". The final positions
are likely to be good to 3" (half a pixel).
All the maps presented here were obtained from the CVF data cube
and have approximately the same resolution: namely, 6" pixels at
the short wavelengths increasing to about 8" at
15 µm ; see Appendix C for more details.
In several of these maps a faint emission can be seen on the
south-east part of the ISOCAM field of view. This feature does not
correspond to anything conspicuous in published images of the region,
in particular in the near-IR images of Marconi et al. (1998). It
is a spurious feature due to multiple reflections of the strong
Trapezium between the detector and the CVF filter wheel, as shown by
the ISOCAM ray tracing studies of Okumura (1999).
The complete SWS scan (2.4-46 µm) was reduced with the
latest version of SWS-IA running at the Institut d'Astrophysique
Spatiale. Calibration files version CAL-030 were used.
Fig. 3 presents the SWS spectrum
( ) obtained inside the
H II region at the position indicated on
Fig. 2. Fig. 3 also shows the comparison of the SWS spectrum
with that of the CAM-CVF pixels averaged in the SWS aperture. The
agreement between these spectra is excellent, well within 20 percent
for the continuum.
![[FIGURE]](img21.gif) |
Fig. 3. The SWS spectrum (full line) compared to the CAM-CVF spectrum (dotted line). In this latter case, all the CAM pixels falling in the SWS aperture have been co-added.
|
© European Southern Observatory (ESO) 2000
Online publication: June 8, 2000
helpdesk@link.springer.de  |