Astron. Astrophys. 363, 1013-1018 (2000)
2. Observations, data reduction and deadtime corrections
For our analysis we used the data from the Proportional Counter
Array (PCA) aboard RXTE. To study the source variability in the hard
spectral state, we chose a compact group of observations of Cyg X-1
performed on 15-17 Dec. 1996 (proposal P10236) with total exposure
time of 80 ksec and 56 available
observations of Cyg X-1 from the proposal P30157 performed between
Dec. 1997 and Dec. 1998 with total exposure time of
110 ksec. The soft spectral state
data were accumulated on 4-18 June 1996 data (proposal P10512, see
also Cui et al. 1997) with total exposure time of
10 ksec. All observations were
performed with 5 PCUs switched on.
The power density spectra were calculated from the light curves
with the bin duration equal to the intrinsic time resolution of the
data
( s s
in proposals P10236 and P10512 and
s s
in proposal P30157). Some deviations from the noise level predicted by
the PCA dead time models are observed in the power spectra constructed
in the PCA energy sub-bands with the amplitude of
(rms/mean)2/Hz. We
therefore followed recommendations of the RXTE GOF and analyzed the
light curves in the total PCA energy band, where these deviations seem
to be considerably less significant (William Zhang, private
communication).
The power spectra calculated for individual time series were
normalized to units of squared fractional rms (Miyamoto et al.
1991)
![[EQUATION]](img26.gif)
and averaged. In the above formula
is the Fourier amplitude at the
frequency ;
- is total observed number of counts
in the time series; - averaged
observed count rate for the time series. Note that the ideal
Poissonian noise component is already subtracted, therefore
for a Poissonian distribution of
counts in the time series.
In a real detector various effects could lead to modification of
the Poissonian noise level by an amount
. Two major effects contribute to
in the case of PCA detector:
![[EQUATION]](img33.gif)
where is a modification of the
noise level due to counts dead time,
- additional noise component resulting from vetoing of the PCA
detectors by the Very Large Events (VLE).
With an accuracy sufficient for our purpose the PCA dead time
caused by the incident photons can be described as a non-paralizable
process 1 (Jahoda
et al. 1996), for which modification of the noise level was derived by
Vikhlinin et al. 1994 (see also Zhang et al. 1995 for the expression
accounting for finite length of the time series). We shall use below
Eq. (A4) from Vikhlinin et al. 1994 transformed to the units of
squared fractional rms:
![[EQUATION]](img36.gif)
where
![[EQUATION]](img37.gif)
Here R is the observed count rate,
is the photon dead time,
is the bin duration, f is the
frequency. For the practical purposes
of the order of 10 is sufficient.
This equation (for ) is valid for an
infinitely long time sequence. One can use more complicated expression
(see Eq. (44) from Zhang et al. 1995) if the analyzed time
sequence contains only small number of bins.
To the first approximation the Very Large Events lead to appearance
of an additional (positive) component (Zhang et al. 1996)
![[EQUATION]](img42.gif)
where - the duration of the VLE
window, - the VLE count rate in one
PCU. Note that the above expression does not include the binning and
sampling effects and is valid in the limit
.
Finally, the total noise component model in the first approximation
would be:
![[EQUATION]](img46.gif)
![[EQUATION]](img47.gif)
where - the total observed
count rate in one PCU, - number of
PCUs. The and
are given by the Eqs. (1) and
(2) respectively. The factor
accounts for the fact that the count streams from
independent units were merged
together. The factor accounts for
reduction of the observed count rate due to Very Large Events. This
formula is valid under the following conditions:
,
and . It is important to stress out
that in the adopted dead time model
is the the total observed count rate of the events causing the
dead time (as opposite to the
observed count rate in the analyzed time series). In the case of PCA
detector it should include total Good Xenon rate, Propane and
Remaining counts. We also note that for sufficiently small count
rates, 3-5 kcnts/s/PCU dependence of
and, therefore, of
upon R vanishes. The model
has been tested on a series of Monte-Carlo simulations and has proven
to be sufficiently accurate in the parameters range of interest.
Note that we did not include in our model an additional background
term (Jahoda 1998 or Jernigan et al. 2000). For bright sources
( 1 kcnts/s/PCA) its contribution can
be neglected.
To verify our model for the noise component we compared it with the
data of PCA observations of an extremely bright source Sco X-1
( 100 000 cnts/s/PCA) for which
all flavors of the dead time distortions are much more prominent than
for Cygnus X-1. We used 3 different observations of Sco X-1-
10059-01-01-00 (Feb. 14, 1996; hereafter #1), 10059-01-03-00 (Feb. 19,
1996; #2) and 10056-01-02-02 (May 26, 1996; #3), having different
value of the VLE window and
different time resolution. Ideally, once the values of the dead time
and VLE window
are calibrated, the model should
reproduce the noise level with count rates
and
set to the measured values. However,
as we discuss below, this is not strictly the case. We therefore
followed the approach adopted by Jernigan et al. 2000 and fitted the
power spectra at high frequencies,
Hz, with the model for the noise component plus the model for the
source component leaving the noise model parameters free. The source
contribution in this particular case was modeled as a superposition of
two Lorentzians representing kHz QPOs. This approach has a
disadvantage that it requires a priory assumptions about the shape of
the power spectrum of the source. In particular, if the source has
very weak and very flat power spectrum component it might be not
detected by our procedure.
In Fig. 1, Fig. 2 and Fig. 3 we present the power
spectra of Sco X-1 in observations #1 #2 and #3 along with the best
fit noise models and residuals. The best fit parameters and their
expected values are given in Table 1. For observation #3 we fixed
parameters and
because the limited frequency range
of the power spectrum ( 4096 Hz) does
not allow us to constrain their values from the fit. From Fig. 1,
Fig. 2 and Fig. 3 one can see that with appropriate tuning
of the parameters the adopted noise model is capable to reproduce the
observed shape of the noise component.
![[FIGURE]](img68.gif) |
Fig. 1. The power spectrum of Sco X-1 from the observation #1 performed with medium VLE window. The solid line shows the best fit model consisting of the noise level model and two kHz QPOs (see text). The dashed line shows the noise level component due to photon dead time. The lower panel shows the residuals data-model.
|
![[FIGURE]](img70.gif) |
Fig. 2. The same as Fig. 1 but for observation #2, performed with short VLE window.
|
![[FIGURE]](img74.gif) |
Fig. 3. The same as Fig. 1 but for observation #3, performed with short VLE window and lower time resolution, s.
|
![[TABLE]](img76.gif)
Table 1. The expected and best fit model parameters of the deadtime correction for 3 observations of Sco X-1.
Notes:
a) e.g. Jahoda et al. 1996, Jahoda et al. 1997, Jahoda 2000
b) measured total count rate (Good Xenon, Propane Counts and Remaining Counts) per PCU
c) measured Good Xenon count rate per PCU
d) preflight values for all PCA, see Giles 1995
e) measured count rate (HouseKeeping data)
In general the best fit parameters of the noise level model
(Table 1) look reasonable and are close to the expected values.
The best fit value of , in
observation #2 is by higher than the
preflight value. It is possible, however, that the duration of the VLE
window differs from the preflight values (Alan Smale, private
communication). The most apparent difference is a large, by a factor
of , discrepancy between the best fit
and observed total count rate. But we should note that the observed
GoodXenon count rate is much closer to the best fit
value (see Table 1). The reason
of this discrepancy is not clear, it might be due to unaccounted
processes in the PCA detectors at high count rates. We note however,
that dependence of the noise component (in units of squared fractional
rms) on the total count rate is a specific feature of a
non-paralyzable dead time process at sufficiently high count rates. At
lower count rate
( 3000-5000 cnts/s) and in
particular in the case of Cyg X-1
( 1000-1500 cnts/s) this
dependence vanishes.
© European Southern Observatory (ESO) 2000
Online publication: December 5, 2000
helpdesk@link.springer.de  |